Page 1 Welch Allyn Connex® ProBP™ 3400 digital blood pressure device Directions for use.; Page 2 Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual. Aug 22, 2017 Welch Allyn Home 1500 Series Upper Arm Blood Pressure Monitor with Easy Bluetooth Smartphone Connectivity Trusted in the doctor’s office, made for the home Trust clinical-grade blood pressure technology from the brand used in 9 out of 10 physician's offices. Welch Allyn's ProBP™ 2400 Digital Blood Pressure Device features SureBP® technology for blood pressure readings in only 15 seconds. The device also provides BP averaging to help ensure accurate hypertension diagnosis. We offer a wide variety of different types of blood pressure monitors and cuffs equipment for your practice. Our products include aneroid sphygmomanometers, digital blood pressure monitors and numerous blood pressure monitor supplies, which may be automatic blood pressure monitors or manual blood pressure monitors. Digital blood pressure. The ProBP™ 2400 digital blood pressure device meets the Canadian Hypertension Education Program (CHEP) guidelines for automated office blood pressure (AOBP) measurement. It is part of a broad portfolio of Welch Allyn devices featuring blood pressure.
Advertising
ProBP3400 error codes
The following tables list the error messages associated with the Welch Allyn ProBP 3400.
Error code
Related System
Additional details
000000001
Persistent storage error
EERPOM is corrupt or invalid
000000001
BP module failure
Mod F NIBP board failure
Error condition
Definition / Description
Required action
Low battery 5 minutes remaining.
The battery is extremely low. Initial
alarm condition occurs when 5 minutes of battery power remain; this is the second low battery alarm.
Click the OK button to dismiss the
message.
Plug the device into the outlet to
dismiss the message.
Low battery 30 minutes remaining.
The battery is extremely low. Initial
alarm condition occurs when 30 minutes of battery power remain; this is the first low battery alarm. Welch Allyn Probp 3400 Calibration
Click the OK button to dismiss the
message.
Plug the device into the outlet to
dismiss the message.
Powering down. Call for service.
This message could be caused by
various internal errors, power manager, and battery faults.
The device will auto power off.
Battery is absent or faulty.
Battery is absent or faulty.
Plug the device into the outlet to
dismiss the message.
User cancelled NIBP reading.
Blood pressure reading cancelled by
user.
Retake measurement.
Inflation too quick; check NIBP cuff
and tubing connections.
The blood pressure cuff inflated too
quickly.
Check for kinked hose.
Cuff pressure limits exceeded.
Powering down. Welch Allyn Probp 3400 Series
Overpressure. Exceeded maximum
limit for this patient mode.
Limit patient movement.
NIBP air leak; check cuff and tubing
connections. Welch Allyn Bp Machine Manual
Major air leak.
Check connections (device will
power down).
Unable to determine NIBP; check
connections; limit patient movement.
Excessive patient movement.
Call Welch Allyn service.
NIBP not functional. Call for service.
Auto zero failure.
Unable to determine NIBP; check
inflation settings.
Could not reach target pressure.
Check for air leaks.
Incorrect NIBP cuff size. Check
patient type.
The patient is outside of the module's
capability to measure NIBP.
Retake measurement.
Inflation too quick. Check NIBP cuff
and tubing connections.
Inflation was too quick; check for a
kinked hose.
Take corrective action; retake
measurement.
Unable to determine NIBP. Check
inflation settings.
Reading couldn’t be obtained
because a high enough inflation couldn’t be reached.
Retake measurement.
OpenTopic | ProBP3400 error codes | 5
Advertising
Popular Brands
Popular manuals
Advertising
1. Connect a USB cable to the service test box; connect the opposite end to a PC.
2. Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box.
3. Connect the blood pressure hose to the NIBP module and the test volume.
•
Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite end to the square
fitting connector.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end to the calibrated
pressure meter.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end to the test volume.
•
Connect the last piece of tubing to the square fitting connector; connect the opposite end to the 'Y' fitting on
the two-tube blood pressure connector.
OpenTopic | ProBP 3400 calibration and verification setup | 9